Guide creation, verification, and maintenance of software system requirements derived from system needs (IEC 62304:2006+A1:2015, 5.2)—with content appropriate to the device, risk controls embedded for Class B/C, and evidence that requirements are testable and traceable. Use when working on software requirements in medical device software.
- Author
- aminalam ↗
- Repository
- meddev-agent-skills ↗
- Licence
- MIT
- Last repository change
- 19.10.2026
Assessment and security
- Files in this package
- 1
- Scripts
- 0
- HTML comments
- 0
- URLs in SKILL.md
- 0
No signals detected. Repository changed on 19.10.2026, without references to retired models or APIs.
MIT, open: use and modify while retaining author attribution.
What it does and when to use it
Guide creation, verification, and maintenance of software system requirements derived from system needs (IEC 62304:2006+A1:2015, 5.2)—with content appropriate to the device, risk controls embedded for Class B/C, and evidence that requirements are testable and traceable. Use when working on software requirements in medical device software.
Original author description
Guide creation, verification, and maintenance of software system requirements derived from system needs (IEC 62304:2006+A1:2015, 5.2)—with content appropriate to the device, risk controls embedded for Class B/C, and evidence that requirements are testable and traceable. Use when working on software requirements in medical device software.
How it works
- Purpose
- When to Apply
- Requirements (testable)
- Recommended Practices
- Patterns
- Anti-Patterns (risks)
- Verification Checklist
- Traceability
- References
- Changelog
SKILL.md preview
---
name: meddev-software-requirements
description: Guide creation, verification, and maintenance of software system requirements derived from system needs (IEC 62304:2006+A1:2015, 5.2)—with content appropriate to the device, risk controls embedded for Class B/C, and evidence that requirements are testable and traceable. Use when working on software requirements in medical device software.
license: MIT
metadata:
skill_id: DOC-SW-REQ
version: 1.0.0
last_updated: '2026-05-21'
applies_to: Class A, Class B, Class C
jurisdiction: FDA, EU MDR, Global
prerequisites: REG-IEC62304, REG-ISO14971
category: documentation
---
# Software Requirements Analysis
## Purpose
Guide creation, verification, and maintenance of software system requirements derived from system needs (IEC 62304:2006+A1:2015, 5.2)—with content appropriate to the device, risk controls embedded for Class B/C, and evidence that requirements are testable and traceable.
## When to Apply
- Defining or updating software requirements before design or implementation.
- Refactoring features that change inputs, outputs, alarms, security, or clinical behaviour.
- Adding SOUP, connectivity, or UI flows that need explicit software requirements.
- Preparing verification plans that must cover all software requirements (5.7).
## Requirements (testable)
1. Derive from System Needs [Class A, B, C]: Document software system requirements from system-level needs; for software-only devices, align system and software requirement sets explicitly. Rationale: 5.2.1.
2. Content Coverage [Class A, B, C]: Address applicable categories—functional capability, inputs/outputs, external interfaces, software-driven alarms/warnings/messages, security, user-interface behaviour, data structures, installation/site constraints, operation/maintenance, IT-network behaviour, user maintenance, and applicable regulatory needs. Rationale: 5.2.2 (interpretation: engineering checklist, not exhaustive copy).
3. Risk Controls in Requirements [Class B, C]: Include software-implemented risk control measures in requirements when identified by risk management. Rationale: 5.2.3.
4. Re-evaluate Device Risk [Class A, B, C]: When software requirements are established or materially changed, re-evaluate device risk analysis and update as needed. Rationale: 5.2.4.
5. Keep System Requirements Current [Class A, B, C]: Re-evaluate and update related system requirements when software requirements change. Rationale: 5.2.5.
6. Requirements Verification [Class A, B, C]: Verify and document that software requirements implement system needs (including risk controls), are non-contradictory, unambiguous, testable, uniquely identified, and traceable to system requirements or other approved sources. Rationale: 5.2.6.
## Recommended Practices
- Use stable IDs (`REQ-62304-###`) in a managed file (e.g. `requirements.yaml`), not only in code comments.
- Add explicit negative, boundary, and security abuse cases per requirement where safety-relevant.
- Tag requirements with safety class of the implementing item when known.
- Cross-reference usability needs (IEC 62366) and cybersecurity expectations in dedicated requirement subsections.
## Patterns
Requirement entry (YAML):
```yaml
requirements:
- id: REQ-62304-102
class: Class C
summary: Stop infusion within 50 ms when door open sensor asserts
sources: [SYS-REQ-44]
risk_controls: [RISK-CTRL-19]
categories: [functional, alarms, timing]
verification: [TEST-201, TEST-410]
negative_cases:
- sensor bounce within 50 ms window
- door open during self-test
```
Security and network excerpt:
```yaml
- id: REQ-62304-210
class: Class B
summary: Reject BLE pairing without LE Secure Connections
categories: [security, interfaces]
verification: [TEST-BLE-03]
```
## Anti-Patterns (risks)
- Requirements only in tickets or chat, not under configuration control -> risk: traceability and audit gaps.
- Vague acceptance criteria ("fast enough", "secure") -> risk: 5.2.6 verification failure.
- Missing risk control requirements for Class B/C -> risk: controls implemented but not verified against needs.
- System requirements not updated when software scope grows -> risk: device-level inconsistency.
## Verification Checklist
- [ ] Software requirements derived from documented system needs.
- [ ] Applicable content categories addressed for the device type.
- [ ] Class B/C: software risk controls reflected in requirements.
- [ ] Device risk analysis re-evaluated after requirement changes.
- [ ] System requirements updated when software requirements change.
- [ ] Verification record shows testability, uniqueness, consistency, and traceability.
## Traceability
- Map `REQ-###` ↔ `SYS-REQ-###` ↔ `HZ-###` / `RISK-CTRL-###` ↔ `TEST-###` ↔ design (`DES-###`).
- See `DOC-TRACEABILITY` for matrix conventions.
## References
- IEC 62304:2006+A1:2015, 5.2 (software requirements analysis).
- ISO 14971:2019 (risk control inputs to requirements).
- `REG-IEC62304-SWRM` (Clause 7 controls in requirements).
- ISO/IEC 25010 (informative quality characteristics for requirement completeness).
## Changelog
- 1.0.0 (2026-05-21): Initial skill for IEC 62304 5.2 software requirements analysis.
Versions
The database has no other skills named meddev-software-requirements.
Database snapshot: 13.11.2026 · package version 1944058e1a03 (SHA-256 digest of the entire package)
Downloads in the last 30 days: 3 · ranking
Report a problem with this skill: harmful content, incorrect descriptions or an author’s removal request.
Similar skills
- 100meddev-software-maintenance
Guide post-release software maintenance per IEC 62304:2006+A1:2015, Clause 6—planning feedback handling, evaluating problems and changes, re-applying development activities when needed, and re-releasing under the same controls as initial release (5.8). Use when working on… - 100meddev-iec-62304-software-risk
Guide implementation of the software-specific risk management process (IEC 62304:2006+A1:2015, Clause 7) alongside ISO 14971:2019—identifying how software items contribute to hazardous situations, defining and verifying software risk controls, and re-analysing risk when software… - 100neuro-soup-inventory
Build the IEC 62304 SOUP (software of unknown provenance) / third-party software inventory and SBOM starting point for neurotech device or analysis software from a code repository (requirements.txt, pyproject.toml, environment.yml, package.json, installed Python packages). Use… - 87medical-device-software-compliance
Reason about SaMD (Software as a Medical Device) regulatory compliance for FDA, EU MDR, and global submissions. Use when the user asks about IEC 62304 safety classification, ISO 14971 risk management, 21 CFR 820/QMSR quality system requirements, Design History File structure,… - 100srs-generation
Generates professional Software Requirements Specification (SRS) documents based on IEEE 830, ISO/IEC/IEEE 29148, and Amazon technical specification standards. This skill activates when the user needs a requirements document, requirements specification, SRS, functional…