Guide post-release software maintenance per IEC 62304:2006+A1:2015, Clause 6—planning feedback handling, evaluating problems and changes, re-applying development activities when needed, and re-releasing under the same controls as initial release (5.8). Use when working on software maintenance in medical device software.

Licence
MIT
Last repository change
19.10.2026

Assessment and security

100/ 100Class: good
Format25 / 25
Frontmatter with name and description, valid name, description up to 1,024 characters, name matching the folder.
Discoverability25 / 25
Description explains when to use the skill, has a reasonable length and uses third person.
Content35 / 35
At least 150 words and at most 500 lines, headings, step lists, examples or code.
Hygiene15 / 15
No placeholders such as TODO, no Windows paths, with a licence.
Riskno signals detected in the scan
Files in this package
1
Scripts
0
HTML comments
0
URLs in SKILL.md
0
Deprecation

No signals detected. Repository changed on 19.10.2026, without references to retired models or APIs.

Licence

MIT, open: use and modify while retaining author attribution.

What it does and when to use it

Guide post-release software maintenance per IEC 62304:2006+A1:2015, Clause 6—planning feedback handling, evaluating problems and changes, re-applying development activities when needed, and re-releasing under the same controls as initial release (5.8). Use when working on software maintenance in medical device software.

Original author description

Guide post-release software maintenance per IEC 62304:2006+A1:2015, Clause 6—planning feedback handling, evaluating problems and changes, re-applying development activities when needed, and re-releasing under the same controls as initial release (5.8). Use when working on software maintenance in medical device software.

How it works

  1. Purpose
  2. When to Apply
  3. Requirements (testable)
  4. Recommended Practices
  5. Patterns
  6. Anti-Patterns (risks)
  7. Verification Checklist
  8. Traceability
  9. References
  10. Changelog
Original SKILL.md headings · 85 lines, 670 words
SKILL.md preview
---
name: meddev-software-maintenance
description: Guide post-release software maintenance per IEC 62304:2006+A1:2015, Clause 6—planning feedback handling, evaluating problems and changes, re-applying development activities when needed, and re-releasing under the same controls as initial release (5.8). Use when working on software maintenance in medical device software.
license: MIT
metadata:
  skill_id: DOC-SW-MAINT
  version: 1.0.0
  last_updated: '2026-05-21'
  applies_to: Class A, Class B, Class C
  jurisdiction: FDA, EU MDR, Global
  prerequisites: REG-IEC62304, DOC-PROBLEM-RES, DOC-SCM
  category: documentation
---

# Software Maintenance Process

## Purpose
Guide post-release software maintenance per IEC 62304:2006+A1:2015, Clause 6—planning feedback handling, evaluating problems and changes, re-applying development activities when needed, and re-releasing under the same controls as initial release (5.8).

## When to Apply
- Software already released for intended use (field, hospital, or patient environment).
- Processing complaints, support tickets, or internal post-release defect reports.
- SOUP upgrades, security patches, obsolescence, or regulatory-driven software updates.
- Coordinating maintenance with postmarket surveillance (`REG-POSTMKT-LABEL`).

## Requirements (testable)
1. Maintenance Plan [Class A, B, C]: Maintain a software maintenance plan covering feedback receipt/documentation/evaluation/resolution/tracking; criteria for what constitutes a problem; use of software risk management; problem resolution process; configuration management for modifications; and evaluation/implementation of SOUP upgrades, fixes, patches, and obsolescence. Rationale: 6.1.
2. Monitor Feedback [Class A, B, C]: Monitor feedback on released software. Rationale: 6.2.1.1.
3. Evaluate Feedback [Class A, B, C]: Document and evaluate feedback to determine if a problem exists; record problems as problem reports including adverse events and specification deviations. Rationale: 6.2.1.2–6.2.1.3.
4. Safety Impact [Class A, B, C]: Evaluate each problem report for effect on safety of released software and whether a change is required. Rationale: 6.2.1.3.
5. Problem Process [Class A, B, C]: Address problem reports via `DOC-PROBLEM-RES`. Rationale: 6.2.2.
6. Change Request Analysis [Class A, B, C]: Beyond Clause 9 analysis, evaluate each change request for impact on organization, released software, and interfacing systems. Rationale: 6.2.3.
7. Approve Changes [Class A, B, C]: Evaluate and approve change requests that modify released software. Rationale: 6.2.4.
8. User and Regulator Communication [Class A, B, C]: Identify approved changes affecting released software; inform users and regulators per local rules when required (see `REG-POSTMKT-LABEL` for reporting detail). Rationale: 6.2.5 (interpretation).
9. Re-apply Development [Class A, B, C]: Identify and perform applicable Clause 5 activities affected by the modification. Rationale: 6.3.1.
10. Re-release [Class A, B, C]: Release modifications per software release requirements (`CICD-RELEASE`, 5.8). Rationale: 6.3.2.
11. Risk on Changes [Class A, B, C]: Perform software risk management for changes per 7.4 (`REG-IEC62304-SWRM`). Rationale: 6.3.1 note / 7.4.

## Recommended Practices
- Single intake queue merging complaints, support, and internal findings with `PR-###` IDs.
- Maintenance plan version-controlled alongside SDP.
- Define SOUP patch SLAs by safety class and exposure.
- Use modification kits (partial updates) only when release process still satisfies 5.8 and configuration baselines.

## Patterns
Maintenance plan excerpt (YAML):
```yaml
maintenance_plan:
  product: InfusionPump-X
  feedback_channels: [complaints, service_portal, security@company.com]
  problem_criteria: |
    Any reproducible deviation from released spec or any safety/security signal
  processes:
    risk: REG-IEC62304-SWRM
    problems: DOC-PROBLEM-RES
    configuration: DOC-SCM
  soup_policy:
    review_cadence: quarterly
    critical_soup: [tiny-rtos, mbedtls]
```

Released change flow:
```markdown
1. Feedback -> PR-2026-120
2. Safety eval -> change required -> CR-220 approved
3. Repeat 5.5–5.7 scope per impact analysis
4. 7.4 risk analysis -> REG-IEC62304-SWRM
5. Release v1.3.1 per CICD-RELEASE / 5.8
```

## Anti-Patterns (risks)
- Field patches without maintenance plan or problem reports -> risk: uncontrolled post-market software.
- Skipping 6.2.3 system/org impact analysis -> risk: hospital integration or QMS breaks.
- SOUP updates without regression or risk re-analysis -> risk: new causes or broken controls.
- Re-release without 5.8 checks -> risk: incomplete verification or wrong build in field.

## Verification Checklist
- [ ] Maintenance plan current and addresses 6.1 elements.
- [ ] Feedback monitored and evaluated; problems recorded as problem reports.
- [ ] Safety impact assessed per problem; changes approved before implementation.
- [ ] Change requests analysed for org/system impact.
- [ ] User/regulator communication planned where required.
- [ ] Applicable Clause 5 activities and 7.4 risk analysis completed for modifications.
- [ ] Re-release satisfies 5.8 / release management checklist.

## Traceability
- Link maintenance records to `PR-###`, `CR-###`, release versions, and postmarket records.
- See `REG-POSTMKT-LABEL` for vigilance and FSCA specifics.

## References
- IEC 62304:2006+A1:2015, Clause 6 (software maintenance process).
- IEC 62304:2006+A1:2015, 5.8 (re-release).
- `DOC-PROBLEM-RES`, `DOC-SCM`, `DOC-CHANGE-CONTROL`, `CICD-RELEASE`.
- `REG-POSTMKT-LABEL` (complaints, vigilance, field actions).

## Changelog
- 1.0.0 (2026-05-21): Initial skill for IEC 62304 Clause 6 software maintenance.

Versions

The database has no other skills named meddev-software-maintenance.

Database snapshot: 13.11.2026 · package version 437b500b09bd (SHA-256 digest of the entire package)

Downloads in the last 30 days: 0 · ranking

Report a problem with this skill: harmful content, incorrect descriptions or an author’s removal request.

Similar skills

  • 100meddev-software-requirements
    Guide creation, verification, and maintenance of software system requirements derived from system needs (IEC 62304:2006+A1:2015, 5.2)—with content appropriate to the device, risk controls embedded for Class B/C, and evidence that requirements are testable and traceable. Use when…
  • 100meddev-iec-62304-software-risk
    Guide implementation of the software-specific risk management process (IEC 62304:2006+A1:2015, Clause 7) alongside ISO 14971:2019—identifying how software items contribute to hazardous situations, defining and verifying software risk controls, and re-analysing risk when software…
  • 87medical-device-software-compliance
    Reason about SaMD (Software as a Medical Device) regulatory compliance for FDA, EU MDR, and global submissions. Use when the user asks about IEC 62304 safety classification, ISO 14971 risk management, 21 CFR 820/QMSR quality system requirements, Design History File structure,…
  • 100neuro-soup-inventory
    Build the IEC 62304 SOUP (software of unknown provenance) / third-party software inventory and SBOM starting point for neurotech device or analysis software from a code repository (requirements.txt, pyproject.toml, environment.yml, package.json, installed Python packages). Use…

Feedback and reports

Your message goes to Bartosz Sokoliński. An email address is needed only for a reply. See the privacy policy.