Define how to record, investigate, resolve, and trend software problems (defects, anomalies, complaints) per IEC 62304:2006+A1:2015, Clause 9—linking safety evaluation, change control, verification closure, and test documentation. Use when working on problem resolution in medical device software.

Licence
MIT
Last repository change
19.10.2026

Assessment and security

100/ 100Class: good
Format25 / 25
Frontmatter with name and description, valid name, description up to 1,024 characters, name matching the folder.
Discoverability25 / 25
Description explains when to use the skill, has a reasonable length and uses third person.
Content35 / 35
At least 150 words and at most 500 lines, headings, step lists, examples or code.
Hygiene15 / 15
No placeholders such as TODO, no Windows paths, with a licence.
Riskno signals detected in the scan
Files in this package
1
Scripts
0
HTML comments
0
URLs in SKILL.md
0
Deprecation

No signals detected. Repository changed on 19.10.2026, without references to retired models or APIs.

Licence

MIT, open: use and modify while retaining author attribution.

What it does and when to use it

Define how to record, investigate, resolve, and trend software problems (defects, anomalies, complaints) per IEC 62304:2006+A1:2015, Clause 9—linking safety evaluation, change control, verification closure, and test documentation. Use when working on problem resolution in medical device software.

Original author description

Define how to record, investigate, resolve, and trend software problems (defects, anomalies, complaints) per IEC 62304:2006+A1:2015, Clause 9—linking safety evaluation, change control, verification closure, and test documentation. Use when working on problem resolution in medical device software.

How it works

  1. Purpose
  2. When to Apply
  3. Requirements (testable)
  4. Recommended Practices
  5. Patterns
  6. Anti-Patterns (risks)
  7. Verification Checklist
  8. Traceability
  9. References
  10. Changelog
Original SKILL.md headings · 79 lines, 614 words
SKILL.md preview
---
name: meddev-problem-resolution
description: Define how to record, investigate, resolve, and trend software problems (defects, anomalies, complaints) per IEC 62304:2006+A1:2015, Clause 9—linking safety evaluation, change control, verification closure, and test documentation. Use when working on problem resolution in medical device software.
license: MIT
metadata:
  skill_id: DOC-PROBLEM-RES
  version: 1.0.0
  last_updated: '2026-05-21'
  applies_to: Class A, Class B, Class C
  jurisdiction: FDA, EU MDR, Global
  prerequisites: REG-IEC62304, REG-IEC62304-SWRM, DOC-CHANGE-CONTROL
  category: documentation
---

# Software Problem Resolution Process

## Purpose
Define how to record, investigate, resolve, and trend software problems (defects, anomalies, complaints) per IEC 62304:2006+A1:2015, Clause 9—linking safety evaluation, change control, verification closure, and test documentation.

## When to Apply
- Defects found in development, verification, manufacturing, or field use.
- Complaints or internal reports referencing software behaviour, crashes, or security issues.
- Anomalies from integration, system, or regression testing (5.6.8, 5.7.2).
- Periodic review of problem trends for quality and safety signals.

## Requirements (testable)
1. Problem Reports [Class A, B, C]: Create a problem report for each detected software problem; include criticality (performance, safety, security), affected products/versions, and context to support resolution. Rationale: 9.1.
2. Investigation [Class A, B, C]: Investigate causes where possible; evaluate relevance to safety using the software risk process; document outcomes; create change request(s) or document rationale for no action. Rationale: 9.2.
3. Stakeholder Communication [Class A, B, C]: Notify relevant parties of significant problems per manufacturer policy (interpretation). Rationale: 9.3.
4. Change Control Link [Class A, B, C]: Implement approved corrections through the change control process (Clause 8.2 / `DOC-CHANGE-CONTROL`). Rationale: 9.4.
5. Records and Risk File [Class A, B, C]: Retain problem reports and resolution records including verification; update risk management file when safety relevance changes. Rationale: 9.5.
6. Trend Analysis [Class A, B, C]: Periodically analyse problem reports for trends (recurring areas, SOUP, modules). Rationale: 9.6.
7. Resolution Verification [Class A, B, C]: Verify fixes—confirm problem closed, trends addressed, changes implemented in correct builds, and no new problems introduced. Rationale: 9.7.
8. Test Documentation on Retest [Class A, B, C]: After changes, test records include results, anomalies, software version tested, environment, tools, date, and tester identity (align with `DOC-TEST-DOCS`). Rationale: 9.8.

## Recommended Practices
- Use IDs `PR-###` for problem reports and link to `CR-###`, `DEF-###`, commits, and `TEST-###`.
- Separate security-critical from safety-critical criticality; escalate per QMS.
- Feed field problems from postmarket processes (`REG-POSTMKT-LABEL`) into the same problem pipeline.
- Re-assess software safety class when investigation shows misclassification (interpretation; see 6.2.2 note in standard).

## Patterns
Problem report (YAML):
```yaml
problem_report:
  id: PR-2026-088
  source: field_complaint
  software_version: 1.2.3
  criticality: safety
  summary: infusion continues 120 ms after door open
  investigation:
    cause: race between ISR and task in pump_control
    safety_evaluation: REG-IEC62304-SWRM 9.2 — links HZ-07
  disposition: change_required
  change_request: CR-201
```

Trend review log:
```markdown
## Problem trend review 2026-Q2
- 4/12 PRs tagged soup:tiny-rtos — escalate SOUP review
- 2/12 PRs door_sensor path — add TEST-415 regression
```

## Anti-Patterns (risks)
- Fixing code without a problem report or change record -> risk: audit trail and CAPA gaps.
- Closing reports without verification evidence -> risk: unresolved safety issues.
- No trend review -> risk: systemic defects undetected.
- Retest records missing version/environment -> risk: non-repeatable evidence (9.8).

## Verification Checklist
- [ ] Each software problem has a problem report with criticality and context.
- [ ] Investigation and safety evaluation documented; change request or no-action rationale recorded.
- [ ] Corrections implemented via approved change control.
- [ ] Resolution records and risk file updates retained.
- [ ] Trend analysis performed per schedule.
- [ ] Closure verification confirms fix and no new regressions.
- [ ] Retest documentation includes required record fields.

## Traceability
- Link `PR-###` ↔ `CR-###` ↔ commits/PRs ↔ `TEST-###` ↔ `HZ-###` / `REQ-###`.
- Integration and system test anomalies must enter this process (5.6.8, 5.7.2).

## References
- IEC 62304:2006+A1:2015, Clause 9 (software problem resolution process).
- IEC 62304:2006+A1:2015, 8.2 (change control; `DOC-SCM`, `DOC-CHANGE-CONTROL`).
- `REG-IEC62304-SWRM` (safety evaluation in 9.2).
- `DOC-TEST-DOCS` (5.7.5, 9.8 test records).

## Changelog
- 1.0.0 (2026-05-21): Initial skill for IEC 62304 Clause 9 problem resolution.

Versions

The database has no other skills named meddev-problem-resolution.

Database snapshot: 13.11.2026 · package version 6eeda175535b (SHA-256 digest of the entire package)

Downloads in the last 30 days: 0 · ranking

Report a problem with this skill: harmful content, incorrect descriptions or an author’s removal request.

Similar skills

  • 100meddev-iec-62304
    Provide actionable guidance to implement IEC 62304:2006+A1:2015 lifecycle processes—development, maintenance, software risk management, configuration management, and problem resolution—with rigor scaled by safety class (A/B/C). This hub skill indexes clause obligations; load…
  • 100meddev-change-control
    Define controlled change-request workflow for software: impact analysis, approvals, verification scope, and regulatory submission flags. Configuration identification and baselines are in DOC-SCM (IEC 62304 Clause 8.1, 8.3); approved implementation and re-verification align with…
  • 87medical-device-software-compliance
    Reason about SaMD (Software as a Medical Device) regulatory compliance for FDA, EU MDR, and global submissions. Use when the user asks about IEC 62304 safety classification, ISO 14971 risk management, 21 CFR 820/QMSR quality system requirements, Design History File structure,…
  • 100neuro-soup-inventory
    Build the IEC 62304 SOUP (software of unknown provenance) / third-party software inventory and SBOM starting point for neurotech device or analysis software from a code repository (requirements.txt, pyproject.toml, environment.yml, package.json, installed Python packages). Use…
  • 100contract-review
    Use when reviewing a contract clause-by-clause from your party's side — buyer/seller/vendor/licensee. Triggers on 'review this contract', 'redline this MSA', 'is this clause a problem'.

Feedback and reports

Your message goes to Bartosz Sokoliński. An email address is needed only for a reply. See the privacy policy.